The salt spray tester uses a salt Solution or an acid salt solution to accelerate corrosion of the material or product under a certain temperature and relative humidity to reproduce the degree of damage suffered by the material or product within a certain period of time. It is widely used in metal material protective layer, mechanical parts, electronic components, auto parts, industrial products, coatings, etc. to accelerate salt spray corrosion test, to evaluate the performance of material protective layer and product against salt spray corrosion. Put a device in front of you, first install the device before you can use the device better.

Shanghai Xiangjie salt spray test machine installation steps:

1. Install the inside of the salt spray test chamber such as: tower spray device, nozzle, funnel, round bar, V-shaped sample holder, etc.

2. The air compressor is ready, and the intake pipe, inlet pipe, and sewage pipe of the equipment are connected (the sewage pipe is connected to the outside).

3. Add water in the saturated barrel and at the bottom of the box and at the seal. Pour the prepared salt water into the salt water tank (the ratio of pure water and industrial salt is 1:19).

4. Check it again, after confirming that there is no problem, close the lid and connect the power supply. After power-on, set the temperature inside the box, the temperature of the saturated barrel and the experiment time on the meter. You can also set according to the experiment requirements. The continuous and periodic spray methods of the equipment. (Note: The temperature of the saturated barrel is set to be 3 degrees higher than the temperature inside the box during the test to ensure that the temperature inside the box is uniform and free of fluctuations.)

5, after all set, click directly to run.

Note on the use of Shanghai Xiangjie salt spray test machine:

1. Open the water inlet valve drain valve and exhaust valve switch

2. Add water to the sink to the pad position.

3. Pour the prepared sodium chloride brine into the brine replenishing tank, that is, automatically fill the brine into the preheating tank in the test chamber, and let the brine flow to the brine preheating tank.