Step 1: Treat the electrolytic cell. Gently shake the electrolytic cell of the trace moisture analyzer a few times, so that the water on the inner wall of the bottle is fully absorbed by the electrolyte, then put the electrolytic cell into the holder, turn on the stirring switch, adjust the stirring speed, and make the liquid surface form. a small vortex
Insert the measuring electrode plug and the electrolytic electrode plug into the corresponding sockets of the main unit and make them in good contact, and turn on the electrolytic switch. Note: In the state of iodine, the green light of the measurement signal indicator and the electrolysis signal indicator are all extinguished, only one red light is on, and the digital display does not count.Step 2: Adjust the balance of the electrolyte.
If the electrolytic cell is still in the state of deep iodine, use 10ul of micro-injector to extract 10ul of distilled water each time, wipe the needle with filter paper, press the start button, and inject it into the cell surface several times through the hole in the middle of the injection cock. In the following, the color change of the electrolyte is observed at any time until the electrolyte turns pale yellow, and the green light of the measurement signal lamp and the electrolysis signal lamp are instructed, and the counter starts counting. (With fresh electrolyte, and the bottle is dry, about 30~50ul of water needs to be injected.) Wait until the counting stops, then stop stirring, pick up the electrolytic cell and gently shake it a few times to make the inner wall of the bottle again. The moisture is absorbed and can be calibrated after the end alarm.
Step 3: Calibrate the trace moisture analyzer.
After the electrolyte is stable, use 0.3 ul micro-injector to extract 0.3 ul of distilled water, wipe the needle with filter paper, press the start button, and inject it below the liquid level. The instrument starts counting and reaches the end point, showing 300±30 ug of water, continuously calibrated 2~ 3 times, the count value is within the required range, indicating that the instrument is accurate.
Gas Explosion-Proof Motor,Explosion Proof Motors,Gas Explosion Proof Motors,Explosion Proof Electric Motor
Yizheng Beide Material Co., Ltd. , https://www.beidevendor.com